Products Description
YexSensor YEX-ACC-Clean-2 Online Self-Cleaning Bracket: Reliable Accessories for Water Quality Monitoring Systems
For multi-parameter water quality monitoring stations, several sensors may operate continuously in water with suspended solids, biological growth or chemical fouling. YexSensor YEX-ACC-Clean-2 Online Self-Cleaning Bracket provides an integrated mounting and automatic cleaning solution for digital sensors, helping contractors improve data continuity and reduce routine manual cleaning work.
Core Technical Advantages: Easier Installation, Lower Maintenance and Stable Field Operation
Automatic Cleaning for Long-Term Monitoring
The built-in motor, position detection circuit and cleaning brush remove contamination from the sensor surface and help prevent microorganism attachment.
Supports Multiple Water Quality Sensors
The bracket supports optional installation of dissolved oxygen, conductivity, turbidity, pH, ORP and other digital sensors for multi-parameter online monitoring.
External Cleaning Expansion
According to field needs, customers can combine jet pipes, water spray pipes, ultrasonic cleaning heads or sodium hypochlorite antifouling devices with the bracket system.
YexSensor YEX-ACC-Clean-2 Online Self-Cleaning Bracket Application Scenarios
YEX-ACC-Clean-2 is designed for water quality monitoring projects that require practical installation accessories, reduced maintenance workload and stable sensor operation.
Integrated Monitoring Stations
Suitable for online stations requiring several water quality parameters in one installation structure.
Municipal Wastewater Projects
Reduces fouling-related maintenance for sensors in treatment process monitoring.
Seawater or Aquaculture Sites
Supports antifouling designs where biological attachment is a recurring issue.
Industrial Water Systems
Helps maintain sensor cleaning in recycled water, cooling water and process water applications.
YexSensor YEX-ACC-Clean-2 Online Self-Cleaning Bracket Technical Performance Parameters
| Technical Item | Parameter Details |
|---|---|
| Brand / Model | YexSensor / YEX-ACC-Clean-2 |
| Product Type | Online self-cleaning bracket |
| Supported Sensors | Up to four digital sensors for five water quality parameters |
| Cleaning Mode | Motor-driven brush cleaning |
| Default Cleaning Interval | 30 minutes |
| Default Cleaning Speed | 3 revolutions each way |
| Operating Temperature | -5 to 50 deg C |
| Power Supply | 12-24V DC |
| Power Consumption | 0.1W @ 12V |
| Cable Specification | 4-core shielded cable, 5 m, customizable |
| Occupied Space | 85 mm x 85 mm x 20 mm |
| Communication | RS485 / Modbus for cleaning parameter setting |
YexSensor YEX-ACC-Clean-2 Online Self-Cleaning Bracket Installation and Electrical Integration Guidelines
Main Installation and Wiring Requirements
Sensor Contact Distance
The brush should just touch the sensor measuring area. Incorrect distance may reduce cleaning performance or damage the measurement surface.
Cable Definition
Red: 12-24V DC; Black: GND; Blue: RS485 A; White: RS485 B. Check wiring sequence carefully before powering on.
Corrosion Protection
Use waterproof and corrosion-resistant cable treatment for long-term immersion, seawater or outdoor exposure.
Maintenance and Commissioning Guide
Routine Maintenance Tasks
Routine Check
Inspect brush movement, motor response, cable sealing and sensor surface condition regularly.
Modbus Settings
Default Modbus format is 9600, n, 8, 1. Cleaning interval range is 6 to 6000 minutes, and cleaning speed range is 0 to 10.
Commissioning Recommendations
Before formal operation, check mechanical installation, cable sealing, sample flow or cleaning action according to the product type. For integration projects, keep a record of sensor model, installation position, wiring definition, communication address and maintenance interval so that future site service can be performed efficiently.
Frequently Asked Questions (FAQ)
Q1: Is YEX-ACC-Clean-2 suitable for system integrators?
A1: Yes. It is designed as an accessory/component for water quality monitoring projects where reliable installation and easier maintenance are important.
Q2: Which sensors can be used with it?
A2: It can be used with YexSensor water quality sensor systems. Final compatibility should be confirmed according to sensor diameter, mounting thread, cable routing and project configuration.
Q3: Does it help reduce maintenance work?
A3: Yes. The product is intended to improve installation consistency and reduce common field problems such as fouling, unstable mounting, difficult inspection or repeated manual adjustment.
Q4: Can it be used in wastewater or surface water projects?
A4: Yes. It is suitable for water treatment, wastewater, surface water, process water and other monitoring projects when installed with proper protection and maintenance planning.
Q5: What should be checked before installation?
A5: Confirm sensor model, installation thread, cable definition, power supply if required, sample flow condition and whether the installation space allows cleaning and calibration.
Q6: Are waterproof measures required?
A6: Yes. For wet, outdoor or submerged applications, all terminals, cable joints and mechanical interfaces should be sealed to prevent water ingress and corrosion.
Q7: Can parameters be adjusted through RS485/Modbus?
A7: For self-cleaning models, cleaning interval and cleaning revolutions/speed can be configured through RS485/Modbus according to the manual. Passive flow cell components do not require communication.
Q8: What information is needed before ordering?
A8: Please provide sensor model, quantity, installation environment, water quality condition, mounting method, cable length and whether automatic cleaning or flow-through installation is required.