How to Keep Turbidity and COD Optical Sensors Accurate in Field Water

2026-07-12

Turbidity and COD sensors are optical instruments. Customers usually notice errors when the optical window is dirty, bubbles remain on the measuring area, or the sensor is too close to the container bottom or wall.

Short Answer

Keep optical windows clean, avoid bubbles, maintain enough distance from container surfaces, and calibrate with the correct zero and slope standards. For COD sensors, calibrate turbidity first, then perform COD calibration.

Turbidity Sensor Calibration Basics

For zero-point calibration, place the sensor vertically in a zero-turbidity solution and keep the probe tip at least 10 cm from the bottom of the beaker. Wait 3 to 5 minutes until the value stabilizes. For slope calibration, use the proper standard solution and again wait for stability.

Shake the standard solution evenly before calibrating. Uneven standards can make the calibration value unreliable.

COD Sensor Calibration Order

When calibrating COD, the turbidity part of the COD sensor should be calibrated first, followed by COD calibration. For COD zero-point calibration, a 5 mg/L COD solution may be used. For slope calibration, a 100 mg/L COD solution may be used.

During COD calibration, the optical paths must be fully submerged by more than 2 cm and must be free of bubbles.

Cleanliness Is Measurement Accuracy

The measurement window and quartz plates must be clean and free of foreign matter. Mud, fingerprints, algae, oil film or air bubbles can all change the optical signal.

Use clean water and a lint-free cloth for gentle cleaning. Do not scratch the window. In field water with high suspended solids, inspect the window more often.

Installation and Field Precautions

Keep turbidity probes at least 100 mm from the bottom and walls during testing. Avoid fluorescent containers or environments that can interfere with optical measurement. If water flow is rapid, fix the COD sensor securely to avoid impact damage.

Never suspend the sensor by the cable. The cable connector must remain secure, and the cable should not be under tension.

Related YexSensor PDF Guides

For calibration commands and model-specific handling notes, see the Water Quality Sensor Calibration and Usage Precautions PDF and the Water Quality Sensor Usage and Maintenance Instructions PDF.

Need Help Selecting or Maintaining a Sensor?

Tell YexSensor your water type, target parameters, installation method, communication interface and maintenance conditions. We can help you choose a suitable sensor model and prepare practical operation guidance for your site.

Gửi yêu cầu
Hãy cho chúng tôi biết yêu cầu của bạn. Chúng ta cùng trao đổi thêm về dự án.
Hãy gửi yêu cầu để chúng tôi đề xuất cảm biến phù hợp nhanh hơn.

Một yêu cầu rõ ràng giúp chúng tôi xác nhận model, phạm vi đo, phương pháp lắp đặt, tín hiệu đầu ra và bảng dữ liệu phù hợp mà không cần gửi email lặp lại.

  • Loại nước: nước uống, nước thải, nước sông, nước nuôi trồng thủy sản, nước chế biến...
  • Các thông số cần đo: pH, ORP, độ đục, oxy hòa tan, độ dẫn điện...
  • Lắp đặt và đầu ra: chìm/đường ống, RS485, 4-20mA, Modbus...
  • Số lượng, mẫu mã mục tiêu, quốc gia giao hàng hoặc tiến độ dự án
Nếu bạn không chắc chắn cảm biến nào phù hợp, hãy mô tả ứng dụng và phương tiện đo của bạn. Nhóm của chúng tôi sẽ giúp chọn mô hình.