How Should pH and ORP Sensors Be Stored, Cleaned and Calibrated?

2026-07-12

pH and ORP sensors are sensitive electrochemical electrodes. Many field problems come from dry storage, damaged glass bulbs, contaminated platinum surfaces, wrong buffer selection or bubbles inside the bulb.

Short Answer

Store pH and ORP electrodes wet in 3 mol/L KCl solution when they will not be used for a long time. Use buffer solutions close to the target measurement range, clean before and after measurement, remove bubbles, and reactivate slow electrodes with the recommended acid and KCl soaking process.

Choose Calibration Buffers Close to the Application

For pH sensors, pH 6.86 is commonly used for zero-point calibration. For acidic applications, pH 4.00 can be used for slope calibration. For alkaline applications, pH 9.18 is more suitable. Choosing a buffer close to the target water improves practical accuracy.

For ORP sensors, standard solutions such as 86 mV, 256 mV or -40 mV may be used depending on the calibration requirement.

Do Not Let the Electrode Dry Out

Long-term dry storage can damage electrode performance and slow down the response. When not used for a long time, place the electrode in a protective cap containing 3 mol/L potassium chloride solution. For pH sensors, do not store the electrode dry.

Avoid prolonged immersion in distilled water. Distilled water can disturb the electrode condition and is not a proper long-term storage solution.

Cleaning and Slow Response Recovery

Clean the electrode with deionized water before and after measurement. For viscous samples, rinse carefully to remove solvents or residues. If a pH or ORP electrode has been used for a long time and becomes slow or inaccurate, soak the probe in 0.1 mol/L HCl solution for 24 hours, then soak it in 3 mol/L KCl solution for 2 hours.

For ORP electrodes with dirt on the platinum surface, toothpaste can be used to gently polish the platinum surface until it becomes shiny. Do this gently and avoid damaging other parts of the sensor.

Handling Tips Customers Often Miss

The glass bulb at the pH electrode tip must not touch hard objects. Any scratch or breakage can make the electrode fail. Before measurement, remove bubbles inside the glass bulb. During measurement, stir the electrode in the sample and then let it stand so the response can stabilize.

Cable protection is also important. Do not tension the cable or let non-waterproofed terminals enter water.

Related YexSensor PDF Guides

For calibration commands and model-specific handling notes, see the Water Quality Sensor Calibration and Usage Precautions PDF and the Water Quality Sensor Usage and Maintenance Instructions PDF.

Need Help Selecting or Maintaining a Sensor?

Tell YexSensor your water type, target parameters, installation method, communication interface and maintenance conditions. We can help you choose a suitable sensor model and prepare practical operation guidance for your site.

Gửi yêu cầu
Hãy cho chúng tôi biết yêu cầu của bạn. Chúng ta cùng trao đổi thêm về dự án.
Hãy gửi yêu cầu để chúng tôi đề xuất cảm biến phù hợp nhanh hơn.

Một yêu cầu rõ ràng giúp chúng tôi xác nhận model, phạm vi đo, phương pháp lắp đặt, tín hiệu đầu ra và bảng dữ liệu phù hợp mà không cần gửi email lặp lại.

  • Loại nước: nước uống, nước thải, nước sông, nước nuôi trồng thủy sản, nước chế biến...
  • Các thông số cần đo: pH, ORP, độ đục, oxy hòa tan, độ dẫn điện...
  • Lắp đặt và đầu ra: chìm/đường ống, RS485, 4-20mA, Modbus...
  • Số lượng, mẫu mã mục tiêu, quốc gia giao hàng hoặc tiến độ dự án
Nếu bạn không chắc chắn cảm biến nào phù hợp, hãy mô tả ứng dụng và phương tiện đo của bạn. Nhóm của chúng tôi sẽ giúp chọn mô hình.